Performance and Blank Components of a Mass Spectrometric System for Routine Measurement of Helium Isotopes and Tritium by the 3He Ingrowth Method

ebook Vorgelegt in der Sitzung vom 1. Juli 1989 von Otto Haxel · Sitzungsberichte der Heidelberger Akademie der Wissenschaften

By Reinhold Bayer

Performance and Blank Components of a Mass Spectrometric System for Routine Measurement of Helium Isotopes and Tritium by the 3He Ingrowth Method

Sign up to save your library

With an OverDrive account, you can save your favorite libraries for at-a-glance information about availability. Find out more about OverDrive accounts.

   Not today

Find this title in Libby, the library reading app by OverDrive.

Download Libby on the App Store Download Libby on Google Play

Search for a digital library with this title

Title found at these libraries:

Library Name Distance
Loading...
For future studies of oceanic circulation it is necessary to develop the tritium measurement via 3He ingrowth into a routine procedure with a high capacity for efficient use. This paper attempts to demonstrate that this can be achieved using a commercial helium isotope mass spectrometer and special procedures for sample preparation, storage for 3He ingrowth, and 3He transfer from the ingrowth containers into the mass spectrometer. This method allows for measurement with a much higher precision and lower detection limit than is possible with counting techniques. Additionally, the parameters and blanks in routine operation of the system are discussed.
Performance and Blank Components of a Mass Spectrometric System for Routine Measurement of Helium Isotopes and Tritium by the 3He Ingrowth Method